Equipment list
Laboratory for Glass and Thermal Insulation
High-temperature rotation viscometer (IJS made with HAAKE RV2 viscosimeter)

Person responsible: Dr. Jakob König
Glass viscosimeter consists of a vertical tube furnace and HAAKE RV2 viscosimeter. Used for determination of viscosities of glasses and other viscous materials.
Technical details
- Temperature range <1500°C
- Sample volume ~ 100 ml
Climatic test chamber (Vötsch VC4018)

Person responsible: Dr. Jakob König
For testing of mineral fibers, electronic components or other materials under extreme climatic conditions (temperature and air humidity).
Technical details
- Temperature range: -45 °C to +180 °C
- Rel. humidity range: 10% to 98%
- RH between +10 °C to +94 °C
- Volume: 190 l
High Temperature Chamber Furnace (Nabertherm, LHT 17A)

Person responsible: Dr. Jakob König
High temperature furnace for sintering and glass melting.
Technical details
- Temperature range 1200 to 1700°C
- Sample size up to 15x15 cm
- Sample amount up to 1 kg
Laboratory for Green Hydrogen
AEM electrolyzer setup (custom made)

Person responsible: Asst. Prof. Dr. Suraj Gupta
An AEM (anion-exchange membrane) single-cell electrolyzer is used to evaluate water electrolysis performance under controlled operating conditions. It enables testing of catalysts, electrodes, membranes, and membrane–electrode assemblies for water electrolysis, while measuring parameters such as cell voltage, current density, efficiency, polarization behavior, durability, and electrochemical losses.
Technical details
- 5 cm² AEM electrolyzer cell with Ti and Ni flow-channel end plates, stainless-steel connectors/endplates
- NovaMea MEA and PTFE gaskets.
- 10 L temperature-controlled water bath, RT–90 °C, 0.1 °C accuracy.
- PID cell temperature control, RT–90 °C, 0.1 °C accuracy.
- Two-channel peristaltic pump, 0–420 mL/min, ±0.5% accuracy
- H₂ sensor with Modbus output and CE certification.
Potentiostat biologic (VSP-300)

Person responsible: Asst. Prof. Dr. Suraj Gupta
The potentiostat is used for electrochemical measurements in various applications, such as water electrolysis and PEC water-splitting. With multi-channel, we can run up to 4 experiments simultaneously. This one comes with two 10A boosters.
Potentiostat: multi-channel (Corrtest Instruments, CS310X)

Person responsible: Asst. Prof. Dr. Suraj Gupta
The potentiostat is used for electrochemical measurements in various applications, such as water electrolysis and PEC water-splitting. With multi-channel, we can run up to 4 experiments simultaneously.
Technical details
- Number of channels: 4
- Potential control range: ±10V
- Potential accuracy: 0.1%×full range±1mV
- Potential resolution:10μV(>100Hz), 3μV(<10Hz)
- Constant current control range: ±1A
- Current accuracy: 0.1%×full range
- Current resolution: 1pA
- Current range: 2nA ~1A
- EIS 10μHz - 1MHz
- AC signal amplitude: 1mV~2500mV
- Operating System requirements: Windows 10 and above
- Weight: 12KG (4 channel)
Potentiostat (Gamry, 1010E)

Person responsible: Asst. Prof. Dr. Suraj Gupta
The potentiostat is used for electrochemical measurements in various applications, such as water electrolysis and PEC water-splitting.
Technical details
- Max Applied Current - ±1 A
- Maximum Applied Potential - ±12 V
- EIS 10 μHz - 2 MHz
- 2, 3, or 4 cell connections
- Current Ranges - 9 (10 nA - 1 A)
- Minimum Current Resolution - 3.3 fA
- Rise Time <1 μs
- Compliance Voltage - ±20 V (@ 1A)
Gas chromatography (Inficon, Micro GC Fusion 2-Module System F0G580W02W02)

Person responsible: Dr. Marjeta Maček Kržmanc
Qualitative and quantitative analysis of compounds in mixed gas samples (hydrocarbons (C1-C8), H2, O2, N2, CO, CO2).
Technical details
- Micro GC Fusion 2-Module System
- Module A : Backflush 1.0 uL, Rt-Molsieve 5A, 0.25mm, 10m, TCD2, Front (heated)
- Module B : Variable volume, Rt-Q-Bond, 0.25mm, 8m, TCD2, Front (heated)
- Carrier gas (options: He, H2, N2, Ar, requested purity 99,999 %)
- Maximal programme column temperature: 250°C
- Maximal sample inlet gas pressure 172 kPa
Flow photoelectrochemical (PEC) cell (custom made)

Person responsible: Asst. Prof. Dr. Suraj Gupta
The PEC cell is used for measurement of photocatalysts in the form of thin film coatings. The cell can be used in PEC mode, where electrical bias can be applied to the photo-electrode, assisting in complex reactions, such as water-splitting.
Technical details
It is a custom made three point measurement cell, with:
- Ag/AgCl reference electrode and
- Pt wire as a counter electrode
- the 5 mm x 5 mm sample is placed on Cu plate, forming a wireless back contact for working electrode
- top side of sample is illuminated and in contact with electrolyte
- the pump is providing the constant flow of the electrolyte
- membrane between counter and working electrode, allowing the separation of produced oxygen and hydrogen
Photocatalytic water-splitting cell

Person responsible: Asst. Prof. Dr. Suraj Gupta
The photocatalytic water-splitting cell is used to perform photo-induced reactions using photocatalyst powders suspended in aqueous solutions.
Technical details
- Jacketed borosilicate glass cell with a total enclosed volume of 210 mL
- Fitted with a removable Quartz window (60 mm dia.) that allows irradiation of simulated solar light on the enclosed solution
- The cell has gas-tight ports for purging inert gases and can be connected directly to a gas chromatograph for continuous gas detection
Solar Simulator (Newport, 66984-300XF-R1)

Person responsible: Asst. Prof. Dr. Suraj Gupta
The solar simulator's primary function is to generate a collimated beam that replicates solar irradiance, enabling its use in applications like photocatalysis where controlled, consistent light exposure is essential.
Technical details
- 300 W Xenon lamp with AM 1.5 G filter
- Beam size: 33 mm with automatic shutter
- Calibrated Si reference cell
- Bandpass filters with centre wavelengths ranging from 350 nm to 800 nm
Laboratory for Pulsed Laser Deposition (PLD)
Scale in the glove box (Mettler Toledo, XPR 105 DUE/M)
Person responsible: Dr. Darinka Primc
Analytical balance for measurements with high precision 0.1 mg.
UHV PLD Glovebox

Person responsible: Jan Žuntar
For storing air sensitive products, samples in nitrogen atmosphere.
Technical details
- Gas concentration measurements, O2 ~ 0-1000 ppm, H2O ~ 0-500 ppm,
- electrochemical O2 sensor resistant to solvents
- Argon working gas
- Automatic control of both antechambers – time controlled cycles
- Large (D = 40 cm) and small (D = 15 cm) antechambers for sample transfer
- Width: 120 cm; height: 90 cm; depth: 80 cm
HV PLD (Demcon TSST, custom made)

Person responsible: Blaž Jaklič
The HV PLD system is used for thin film growth of functional oxides used in energy storage and conversion devices. HV PLD system is vacuum-connected to a glove box that allows storage of reactive targets and transfer of thin films that are sensitive to ambient atosphere. This PLD system is equipped with reflective high-energy electron diffraction (RHEED) system, which enables the monitoring of thin layer growth and structural analysis of the surfaces of prepared samples. Resistive heater can be used for heating up the substrates up to 900°C. For the ablation of targets, additional optical elements are used for guiding the laser light from the existing KrF pulsed laser.
Technical details
- Temperature range RT to 900 °C with resistive heating
- Pressure range: 10-8 mbar - 1000 mbar
- RHEED gun for in-situ monitoring of surface structure
- oxygen, nitrogen, argon gas inlets
HV PLD Glovebox (MBRAUN, UNIlab CS2)

Person responsible: Jan Žuntar
For sample manipulation and storage under a controlled inert atmosphere (H2O and O2 levels < 1 ppm).
Technical details
- Gas concentration measurements, O2 ~ 0-1000 ppm, H2O ~ 0-500 ppm
- electrochemical O2 sensor resistant to solvents
- Argon working gas
- Automatic control of both antechambers – time controlled cycles
- Large (D = 40 cm) and small (D = 15 cm) antechambers for sample transfer
- Width: 120 cm; height: 90 cm; depth: 80 cm
UHV PLD (TSST, custom made)

Person responsible: Lucija Bučar
The UHV PLD system is specifically used for oxide thin film integration on semiconductor substrates, like Si and Ge. The system is equipped with an ion-getter pump, titanium sublimation pump, valves with metal seals and a bake-out tent that help to maintain an ultra-high vacuum environment (5 × 10-10 mbar). UHV PLD system is vacuum-connected to a glove box that allows storage of reactive targets, a storage chamber for storing up to 12 samples, and an analytical system for temperature-programmed desorption (TPD). This PLD system can also be connected to a vacuum suitcase, which enables the transfer of samples to devices for structural and chemical analysis (XPS, JT-STM) located within other departments of our Institute. It is equipped with reflective high-energy electron diffraction (RHEED) system, which enables the monitoring of thin layer growth and structural analysis of the surfaces of prepared samples. Apart from the resistive heater, a system for IR laser heating is also available enabling substrate heating up to 1200 °C. For the ablation of targets, additional optical elements can be used for guiding the laser light from the existing KrF pulsed laser.
Technical details
- Temperature range +25 to +1200 °C with infra-red laser, also possible resistive heating
- Pressure range: 10-8 mbar - 1000 mbar
- Residual gas analyzer
- Oxygen, nitrogen, argon gas inlets
- in situ transfer via UHV suitcase
Laboratory for Thermal Analysis
Thermal analysis (NETZSCH; STA 449 C Jupiter & QMS 403 Aëolos)

Person responsible: Lucija Bučar
Simultaneous monitoring of mass changes, thermal effects, and the analysis of evolved gases during controlled heating. The STA 449 Jupiter with the TG/DSC or TG/DTA sample holder is used for the study of thermal decompositions, phase transitions, crystallisation, determination of enthalpy and the kinetics of phase transitions.
Technical details
- High temperature tube furnace: - with SiC heater (temperature range 25 °C-1550°C)
- Weighing resolution: 0.1 μg
- Atmosphere: oxidising (Ar/O2,O2), inert (Ar)
- Evolved gas analysis: 1-300 amu
- Sample holders: TG/DSC, TG/DTA, TG
- Crucibles: Pt/Rh, Al2O3
*Financed by Center of Excellence Nanoscience and Nanotechnology